Table 2

Results of remote biomonitoring device testing of newborn baby temperature measurement in different temperature ranges by different groups of neonates

Measurements:
Neonates’ categories
Time of temperature recording (hh:mm)Readings (n)Temperature
Mean (±SD)
Temperature
Mean difference
(first reading)(last reading)Device
(abdomen)
Digital (axilla)(device-digital)
Well babies (in postnatal ward) 
 103:5404: 541336.6136.60.01
 204:0205: 021337.0036.80.18
 304:0505: 051337.0036.80.21
 Subgroup3936.87 (±0.2)36.73 (±0.1)0.13*
Device
(abdomen)
Probe (abdomen)(device-probe)
Sick neonates in NICU under radiant warmer with temperature >35.5°C
 415:4316:431336.6236.570.05
 511:1612:161335.7836.56−0.78
 612:4713:471335.6636.34−0.68
 711:5612:561335.6136.64−1.03
 814:2615:261335.5036.35−0.85
 911:2512:251335.4136.32−0.91
 1014:0015:001335.7636.43−0.67
 1111:0412:041336.7136.450.26
 1212:1713:171335.8436.43−0.59
 1312:3913:391335.8936.47−0.58
 Subgroup13035.88 (±0.4)36.46 (±0.1)−0.6†
Sick neonate in NICU 
 Undergoing therapeutic cooling for birth asphyxia (temperature approximately 33°C)
  Neonate 1402:5103:511333.1033.150.05
 Undergoing rewarming after therapeutic cooling for birth asphyxia (temperature approximately 34°C)
  Neonate 141233.9533.920.03
 Subgroup2533.53 (±-0.6)33.54 (±0.5)0.01‡
Overall15535.76 (±1.1)36.12 (±1.1)−0.4§
  • *t=2.2; df=2; P=0.2.

  • †t=4.4; df=9; P<0.001.

  • ‡t=0.25; df=1; P=0.8.

  • §t=3.0; df=14; P<0.01.

  • NICU, neonatal intensive care unit.