Table 2

Statistical analysis for determination of performance of new rapid test

Total number of samples N=426Kirby-Bauer disc diffusion
UTI positiveUTI negativeTotal
Rapid test
UTI positive2286234
UTI negative15177192
Total243183426
Sensitivity93.82%
Specificity96.72%
PPV97.43%
NPV92.18%
κ0.9
  • NPV, negative predictive value; PPV, positive predictive value; UTI, urinary tract infection.