RT Journal Article SR Electronic T1 Comparative evaluation of a low cost direct ophthalmoscope (Arclight) for red reflex assessment among healthcare workers in Malawi JF BMJ Innovations JO BMJ Innov FD All India Institute of Medical Sciences SP 113 OP 116 DO 10.1136/bmjinnov-2019-000361 VO 6 IS 3 A1 Dooley, Eleanor A1 Kousha, Obaid A1 Msosa, Joseph A1 Ndaule, Eric A1 Abraham, Carl A1 Parr, Jennifer A1 O'Hare, Bernadette A1 Blaikie, Andrew YR 2020 UL http://innovations.bmj.com/content/6/3/113.abstract AB We compared the diagnostic performance and ease of use of a new solar powered low-cost Arclight direct ophthalmoscope (AO) to a more expensive traditional direct ophthalmoscope (TDO) (Keeler Professional V.2.8) in detecting abnormal red reflexes in simulated eyes. Both devices were used by 19 optometry students and 17 paediatric doctors based in the Kamuzu Central Hospital Campus in Lilongwe, Malawi. Participants examined four normal and four abnormal red reflexes using the two devices in random order. We scored the participants on their ability to identify clinical signs and make a diagnosis. Participants scored each device for ‘ease of use’. There was no statistically significant difference in diagnostic performance or ‘ease of use’ between the AO and the TDO when attempting to detect abnormal red reflexes in simulated eyes. We conclude that AO is an inexpensive yet equally effective alternative to the TDO in detecting red reflexes and due to its low cost, portability and consumable independence is well suited for use in low-income and middle-income countries.