RT Journal Article SR Electronic T1 Remote biomonitoring of temperatures in mothers and newborns: design, development and testing of a wearable sensor device in a tertiary-care hospital in southern India JF BMJ Innovations JO BMJ Innov FD All India Institute of Medical Sciences SP 60 OP 67 DO 10.1136/bmjinnov-2016-000153 VO 4 IS 2 A1 Prem K Mony A1 Prashanth Thankachan A1 Swarnarekha Bhat A1 Suman Rao A1 Maryann Washington A1 Sumi Antony A1 Annamma Thomas A1 Sheela C Nagarajarao A1 Hiteshwar Rao A1 Bharadwaj Amrutur YR 2018 UL http://innovations.bmj.com/content/4/2/60.abstract AB Objective Newer technologies such as wearables, sensors, mobile telephony and computing offer opportunities to monitor vital physiological parameters and tackle healthcare problems, thereby improving access and quality of care. We describe the design, development and testing of a wearable sensor device for remote biomonitoring of body temperatures in mothers and newborns in southern India.Methods Based on client needs and technological requirements, a wearable sensor device was designed and developed using principles of ‘social innovation’ design. The device underwent multiple iterations in product design and engineering based on user feedback, and then following preclinical testing, a techno-feasibility study and clinical trial were undertaken in a tertiary-care teaching hospital in Bangalore, India. Clinical trial phases I and IIa for evaluation of safety and efficacy were undertaken in the following sequence: 7 healthy adult volunteers; 18 healthy mothers; 3 healthy babies; 10 stable babies in the neonatal care intensive unit and 1 baby with morbidities. Time-stamped skin temperature readings obtained at 5 min intervals over a 1-hour period from the device secured on upper arms of mothers and abdomen of neonates were compared against readings from thermometers used routinely in clinical practice.Results Devices were comfortably secured on to adults and neonates, and data were efficiently transmitted via the gateway device for secure storage and retrieval for analysis. The mean skin temperatures in mothers were lower than the axillary temperatures by 2°C; and in newborns, there was a precision of –0.5°C relative to axillary measurements. While occasional minimal adverse events were noted in healthy volunteers, no adverse events were noted in mothers or neonates.Conclusions This proof-of-concept study shows that this device is promising in terms of feasibility, safety and accuracy (with appropriate calibration) with potential for further refinements in device accuracy and pursuit of further phases of clinical research for improved maternal and neonatal health.