Description and evaluation of a self-operated waist measurement device

Abstract

Background National guidelines recommend that waist circumference (WC) be measured in patients with a body mass index (BMI) 27–35 kg/m2. Unfortunately, perhaps due to logistical reasons, WC is seldom measured in clinical settings. Herein, we describe the performance of a self-operated waist measurement device (SOWMD) as a potential means to overcome barriers to measuring WC.

Materials and methods Ten volunteers underwent WC measures by professionals and SOWMD on 5 separate days to assess the reproducibility and accuracy. We then compared SOWMD measures with CT-derived fat content by recruiting 81 patients scheduled for a diagnostic abdominal CT scan.

Results There was no difference between professionally measured and SOWMD-measured WC; the intraindividual coefficient of variation over the 5 days was between 0.4% and 2.2%. The WC measured manually, by SOWMD and CT scan were highly correlated (r=0.90–0.92, all p<0.001). The minimal WC measured by SOWMD was a better predictor (r=0.81 for all patients, r=0.87 for men, both p<0.001) of CT-measured visceral adipose tissue volume than other approaches. The minimal WC measured by SOWMD was correlated with fasting plasma glucose (r=0.40, p<0.05), triglyceride (r=0.41, p<0.01) and high-density cholesterol (r=−0.49, p<0.001) concentrations.

Conclusion SOWMD is a reproducible, accurate and convenient way to measure WC that can provide added value for health care providers when combined with BMI information.

  • Diabetes Mellitus
  • Nutritional Sciences
  • Patient Care
  • Primary Healthcare

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.